Construction of lead (II) ion detection method based on regulation of thiacyanine self-assembly by nucleic acid aptamer
CSTR:
Author:
Affiliation:

(1. National Center for Occupational Safety and Health, NHC, Beijing 102308, China;2. Taizhou Medical New & Hi-tech Industrial Development Zone, Taizhou, Jiangsu 225300, China)

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    In this paper, regulation of self-assembly properties of thiacyanine dye (Dye2) by nucleic acid aptamer (T30695) was used to construct a specific recognition unit for detecting Pb2+. G-quadruplexes were formed to interact with dye 2 monomers, which could disassemble dimers of Dye2 into monomer, in the presence of Pb2+. The absorption vales at 422 nm via 445 nm was used as a specific recognition signal to achieve the purpose of specificity and high sensitivity identification of Pb2+ only by UV-vis spectra. This system has a good linear relationship from 0.25 to 10.00 μmol/L (R2 = 0.992 7), and the detection limit is 0.115 μmol/L.